性能特征
Process thermostat for pressurised operation with integrated cooling system for dynamic temperature control within external circuits
Clear text menu navigation, six selectable languages DE, EN, FR, ES, IT, RU
SelfCheck Assistant for system diagnosis
PowerAdapt system for the use of the maximum possible amount of heat permitted by the power supply system
USB and Ethernet interface equipped as standard
Option for upgrading up to 2 additional interfaces (RS 232/485, Profibus, analogue, contact or EtherCAT module)
Digital display of pump pressure
Integrated web server for browser based operation in local area networks via PC, tablet or smart phone, secure data transfer due to authentication procedure and encryption
Utilises traditional refrigerants (HFCs) in accordance with European legislation to control F-gases (EU) 573/2024
For use with non-flammable liquids
Management of heat transfer liquids with stored properties
Fully electronic continuous controller with PID action for internal & external control
Low-level and adjustable over-temperature protection with acoustic alarm
Port for external Pt100 integrated, second external Pt100 feasible via interface module
Integrated and adjustable bypass
Very small internal volume and big non-thermostated expansion vessel (cold fluid layer system)
SmartCool system for energy-saving digital cooling management including compressor on-off control
Coloured TFT display for simultaneous indication of actual & set values and graphic illustration of the temperature profile
Easy input via cursor and soft keys. Additional Tmax key for overtemperature protection
Self adapt function for determination of control parameters
Powerful LAUDA Variopump (pressure pump) with 5 selectable output levels or control of outflow pressure
Remote fault indication through floating contact
Programmer with 150 temperature/time segments that can be separated into 5 programs
Filling by using compressed air or a drum pump
Condenser cooling Water
工作范围
最小工作温度
-40 °C
最大工作温度
140 °C
温度稳定性
0.1 ± K
技术特征
工作温度范围
-40 ... 140 °C
环境温度范围
5 ... 40 °C
温度稳定性
0.1 ± K
最大加热功率
24 kW
最大功率消耗
37 kW
电流消耗
60 A
最大输送压力
6,0 bar
泵最大输送流量(压力)
100 L/min
输入/输出连接螺纹(外部)
M38 x 1,5
输入/输出软管尺寸
1″
最小灌注容积
12.1 L
最大灌注容积
48.1 L
水冷却装置连接螺纹(内部)
3/4 ″
建议冷却水温度
15 °C
Cooling Water Temperature Max
30 °C
冷却水消耗
29 L/min
最小冷却水压差
0.8 bar
Maximum_Pressure_Cooling_Water
10 bar
尺寸(宽x深x高)
1100 x 895 x 1865 mm
重量
625 kg
声压级
74 dB(A)
电源供应
400 V; 3/PE; 50 Hz & 460 V; 3/PE; 60 Hz
电源插头
无电源线的电气连接箱,无插头
温度 | 泵级 | 回火介质 | 冷却能力 50 赫兹 | 冷却能力 60 赫兹 |
---|---|---|---|---|
140 °C | 8 | 水-乙二醇 | 25 kW | 28 kW |
100 °C | 8 | 水-乙二醇 | 25 kW | 28 kW |
20 °C | 8 | 水-乙二醇 | 25 kW | 28 kW |
10 °C | 8 | 水-乙二醇 | 25 kW | 28 kW |
0 °C | 8 | 水-乙二醇 | 24.5 kW | 27.5 kW |
-10 °C | 8 | 水-乙二醇 | 24 kW | 27 kW |
-20 °C | 8 | 水-乙二醇 | 17.7 kW | 19 kW |
-30 °C | 8 | 水-乙二醇 | 11 kW | 12 kW |
-40 °C | 8 | 水-乙二醇 | 7.5 kW | 7.7 kW |
下载 - 一般产品数据
设备 | 设备类别 | 文件类型 | 语言 | 格式 | 尺寸 | |
---|---|---|---|---|---|---|
Integral P | 工艺过程温控器 | 产品说明手册 |
中文
|
3,09 MB | 下载文档 |
适用配件
如需合适的配件,请使用 LAUDA 配件搜索引擎。