LAUDA filling and flow control system

Based on FD 50 and MID 80, 100-240 V; 50/60 Hz

Flow control system with magnetic inductive measurement principle

Flow control range: 0.2 ... 70 L/min

订购编号: L004058

性能特征

Flow control system with magnetic inductive measurement principle
User-interface allows setting and control of the temperature control unit
The maximum controllable volume flow depends on the pump output of the temperature control unit and the pressure drop in the application
Active filling and draining system for temperature control circuits with changing applications
Demand-related manual override of automation possible
Auxiliary pump for active support of filling via the thermostat
Mesh filter against aerosols
Movable on wheels
Permitted heat transfer liquids: water/glycol mixtures
Use in pressurized temperature control units up to max. 140 °C
Adjustable pressure limit in application circuit
Setting the parameters via the operating menu of the temperature control unit
Calibratable pressure sensors
Monitoring of the liquid temperature before compressed air discharge
Communication to the temperature control unit via LiBus
Available accessories: Inline - filter (strainer); Sight glass for venting control; Automatic refill incl. pump; Automatic refill via solenoid valve
Useable on LAUDA Integral variants IN XT and IN P
Use in open or cold-oil overlay systems up to max. 90 °C
With certificate of calibration
Automatic filling and emptying sequence, status visualized via illuminated keys
Pneumatic leak test before filling
Large tank reservoir with level detection for handling large quantities of liquid
Full process integration via connectivity temperature control unit

技术特征(依据 DIN 12876)

工作温度范围
10 ... 40 °C
环境温度范围
5 ... 40 °C
体积流量控制精度(20 °C; 20 升/分钟; 1 巴)
0.2 ± L/min
输入/输出连接螺纹(外部)
M38 x 1,5
尺寸(宽x深x高)
627 x 1033 x 1278 mm
重量
150 kg
最大功率消耗
0.3 kW
电源供应
100-240 V; 50/60 Hz
电源插头
带插头的电源线 (GB2099, 15934)

您的比较清单中已有三种产品。
要比较其他产品,请删除比较清单中的某个产品。