LAUDA flow control MID 80

LiBus, compatible with Integral IN XT, IN P

Flow control system with magnetic inductive measurement principle

Flow control range: 0.2 ... 70 L/min

Order no.: L003217

Features

Flow control system with magnetic inductive measurement principle
Use in pressurized temperature control units up to max. 140 °C
Adjustable pressure limit in application circuit
Permitted heat transfer liquids: water/glycol mixtures
Use in open or cold-oil overlay systems up to max. 90 °C
With certificate of calibration
User-interface allows setting and control of the temperature control unit
The maximum controllable volume flow depends on the pump output of the temperature control unit and the pressure drop in the application
Movable on wheels

Technical data (according to DIN 12876)

Working temperature range
-40 ... 140 °C
Ambient temperature range
5 ... 40 °C
Flow control range
0.2 ... 70 L/min
Flow measuring range
0.0 ... 99 L/min
Control accuracy volume flow (20 °C; 20 L/min; 1 bar)
0.2 ± L/min
Measurement deviation
± 3,3 % at 1 L/min; ± 0,2 % at 70 L/min
In / Outlet connection thread (outside)
M38 x 1,5
Dimensions (WxDxH)
605 x 420 x 620 mm
Weight
59 kg
Max. power consumption
0.02 kW
Power supply
LiBus
Power plug
---